Proving an integration works before Production.

On this page

Validation

Validation exercises the configured interfaces and records the outcome. Production readiness depends on validation evidence, not on a self-declared status.

  • A connection test proves reachability, authentication and contract version agreement — not that the laboratory system accepts orders.
  • Sandbox scenarios prove the full order, specimen, result and exception path against synthetic data.
  • Validation evidence is scoped to one environment: a Sandbox run never counts toward Production, and a stale or failed read is treated as not validated.
  • Any unresolved requirement remains visible as a production blocker until it is genuinely resolved.
  • Native interface validation against a real laboratory system requires that laboratory's approved test endpoint, test patients, signed test plan and clinical sign-off; synthetic tests are never presented as vendor certification.

Owner and IT validation sequence

  • Confirm workspace, selected environment, adapter contract/version and responsible laboratory IT owner. Record the current connection or mapping revision without exporting credentials.
  • Follow the existing six steps: Laboratory information, Testing capabilities, Fulfillment, Test your setup, Connect your laboratory, Production readiness. Identify the LIS product/version and licensed interface in Step 5; Step 6 remains a separate technical and commercial decision.
  • Preflight the approved endpoint/network, catalog orderables, specimen and ordering-party requirements, current immutable profile revision and independent mapping reviewer. Missing tenant scope or installation authority leaves testing blocked; a gateway or Edge probe supplies no approval.
  • Run the supported connection check and inspect its actual outcome. Reachability and authentication are prerequisites, not order acceptance.
  • In Sandbox, create only synthetic scenarios through the existing workflow. Correlate the order, acknowledgement, specimen/result and exception records in that same environment.
  • Resolve missing mappings, unsupported collection/region requirements and exceptions through their canonical workflows, then repeat the affected synthetic scenario. A display label is not evidence of resolution.
  • Inspect current evidence after refresh; failed, missing, expired or wrong-path evidence remains incomplete. Step 5 completion and Step 6 readiness are separate conclusions.
  • Laboratory connection/configuration changes and deployment validation require authorized Lab Admin or platform-operator actions. Operations contributes IT information, not self-granted Production approval.

Four different test boundaries

  • Synthetic Sandbox: the existing fixture journey demonstrates its configured synthetic path. It never certifies a real LIS or counts as Production validation.
  • Standalone protocol tests: locally hosted REST/FHIR, HL7/MLLP and SFTP peers check parsing, envelopes and failure handling in source. They do not demonstrate Core installation, hosted dispatch or vendor compatibility.
  • Approved real laboratory test: requires the laboratory's explicit endpoint/test identity authorization, signed test plan, exact profile/mapping revision and independent clinical review. No real-system test is established for the candidate vendors.
  • Post-live Production diagnostics: fresh, authorized metadata-only reads inspect current credentials/grants, connection and activity state. They are not permission to create a clinical order, inject a result, use a synthetic patient or rerun Sandbox in Production.

Evidence correlation, expiry and clinical sign-off

  • Retain protocol, authorized profile revision, reviewer and proof source alongside safe correlation/idempotency references and the observed test date.
  • Reconcile accession/specimen, supported orderables, observation codes/units and partial/final/corrected/amended versions with authorized release status.
  • Negative cases include timeouts, lost acknowledgement, wrong environment, revoked credentials, conflicting mappings and duplicate source events. Refusal or reconciliation is not a completed pass.

Keep the safe run/request reference, workspace, environment, selected path, adapter/profile/mapping revision, observed timestamps and order/result correlation together. Compare any expiry supplied by the canonical evidence; do not invent a universal validity duration. Replaced configuration, an expired approval or a failed current read cannot inherit an old pass.

Transport receipt, HL7 commit acknowledgement, application acceptance, accession/specimen acceptance and final or corrected result release are separate milestones. The laboratory reviewer confirms clinical mapping, units, status semantics and release policy; Solia does not infer sign-off from a successful request.

Closed gates and escalation

Important

Production promotion remains paused and native execution remains disabled. A successful local test, payment, API key, owner role or completed checklist cannot open those gates.

When validation is refused, preserve its safe error and current evidence references, identify the missing owner/approval and escalate through the existing integration workflow. Do not use a generic clinical write or a manual resend to force a pass. Authorized Production, first real LIS certification and second-site zero-source-change reuse remain separate blocked acceptance milestones.

Selected evidence scope

The selected CI verifies 29 Prompt4 candidate wire/draft tests, separate disposable two-buyer 7/7 and integrated 24/24 within 1,632 selected tests. The candidate callback is not one joined canonical transaction. Distinguish pure transformation, loopback, disposable persistence and separately hosted verified acceptance.